High and Low Temperature (Humidity) Test Chamber — ETH Series

TestEQ ETH-series temperature and humidity test chambers reproduce controlled temperature and humidity conditions for component, assembly and material reliability testing. The series uses application-specific compressors from 0.8 to 8 HP and an optimized air circulation system.

High and low temperature humidity test chamber

Applications

  • Electronics and PCB assemblies: damp heat, temperature and humidity exposure
  • Semiconductor and electronic components
  • Automotive and EV components
  • Aerospace and defense equipment
  • Batteries and power electronics
  • Materials, coatings and adhesives

Chamber configuration

  • Cabinet models from 245 L to 1000 L; large walk-in temperature-humidity systems can be customised up to 4500 m³.
  • Temperature range -40 °C to +180 °C or -70 °C to +180 °C, depending on model.
  • Humidity range 10 % to 98 % RH (5 % to 98 % RH optional), with pump water supply.
  • Constant temperature control ≤ ±0.3 °C for testing stability.
  • Dew point temperature range 7 °C to 87 °C; extended dew point conditions -10 °C.
  • Application-specific compressors from 0.8 to 8 HP deliver fast product temperature change rates; faster transitions shorten the test cycle.
  • Optimised air circulation improves temperature gradient, control precision and reliability.
  • 7" colour touchscreen controller with program mode (20 programs, up to 50 steps each, up to 20 loops) and fixed-value mode; RS-485 local and remote communication, up to 16 chambers connected with a maximum total cable length of 800 m.
  • Over-temperature protection, power-failure protection, upper/lower temperature limits, timed start/stop and self-diagnosis.
  • Noise ≤ 68 dB.

Humidity chamber interior

Specifications (ETH series)

ParameterETH225-40ETH225-70ETH408-40ETH408-70ETH1000-40ETH1000-70
Capacity (m³ / L)0.245 / 2450.245 / 2450.408 / 4080.408 / 4081.0 / 10001.0 / 1000
Interior W×H×D (mm)700×700×500700×700×500700×800×730700×800×7301000×1000×10001000×1000×1000
Exterior W×H×D (mm)900×1680×1300900×1680×1300900×1780×1500900×1780×15001200×1980×17001200×1980×1700
Temperature range-40 to +180 °C-70 to +180 °C-40 to +180 °C-70 to +180 °C-40 to +180 °C-70 to +180 °C
Temperature fluctuation≤ ±0.5 °C≤ ±0.5 °C≤ ±0.5 °C≤ ±0.5 °C≤ ±0.5 °C≤ ±0.5 °C
Temperature deviation≤ ±2.0 °C≤ ±2.0 °C≤ ±2.0 °C≤ ±2.0 °C≤ ±2.0 °C≤ ±2.0 °C
Temperature uniformity≤ 2.0 °C≤ 2.0 °C≤ 2.0 °C≤ 2.0 °C≤ 2.0 °C≤ 2.0 °C
Cooling +25 → -20 °C (min)301830183019
Cooling +25 → -40 °C (min)603560326032
Cooling +25 → -70 °C (min)—80—80—80
Heating -25 → +150 °C (min)565656565656
Heating -40 → +85 °C (min)505050505050
Heating -70 → +85 °C (min)—60—60—60
Humidity range (%RH)10–98 %RH10–98 %RH10–98 %RH10–98 %RH10–98 %RH10–98 %RH
Humidity uniformity±1.0 %RH±1.0 %RH±1.0 %RH±1.0 %RH±1.0 %RH±1.0 %RH
Relative humidity deviation≤ +2.0 to −3 %RH (>75 %RH) / ≤ ±5.0 %RH (≤75 %RH)as leftas leftas leftas leftas left
Rated current (A)121514212538
Rated power (kW)687101218
Power supplyAC380V, 3 phase 5 lines, 50/60 Hz

Cooling/heating times are for an empty chamber. The humidity indicator applies to humidity-configured chambers; 5 % to 98 % RH is an optional condition. High-humidity options include 85 °C / 85 %RH for 1000 hours at least, 40 °C / 75 %RH, 55 °C / 10 %RH and 65 °C / 95 %RH. Performance is measured at 50 Hz and 23.9 °C laboratory conditions. These chambers are not intended for flammable, explosive or corrosive test items, or for biological samples.

Touchscreen controller

Standards

Chambers can be configured for test programs based on IEC 60068, JEDEC JESD22, MIL-STD-810 and ISO 16750 requirements.

For a chamber configured to a specific temperature range, humidity condition or test standard, contact us with the specimen dimensions and required profile. Related: Thermal Shock Test Chamber , Rapid Temperature Change Chamber .

Tell us the test standard and the specimen

Send the standard, the specimen size and heat load, the temperature or pressure range. Our engineers will come back with a chamber configuration.